An independent 2026 study in Frontiers in Human Neuroscience (the link is at the bottom of this post) recorded auditory ERPs with a semi-dry (saline-sponge) Smarting mobile EEG system and a wired, gel-based 64-channel BioSemi lab system in the same shielded laboratory. Using spatial clustering, Smarting detected individual-level N100 and P300 responses in 34 of 36 subject-level tests, compared with 34 of 40 for BioSemi (n = 9 and n = 10 participants). The authors identify Smarting as one of two research-grade mobile systems whose signal quality their results validate.
Why do individual-level ERPs matter?
Most ERP research reports group averages. Precision EEG aims to use ERPs as markers for individual people, for example clinical assessment, development tracking or personalised interventions. That requires each participant’s response to be detectable and reliable, not just the group mean.
Korisky and colleagues asked whether classic auditory ERPs can be measured reliably at the individual level, and whether that holds across different EEG systems, including mobile ones.
How was the study designed?
Participants completed a three-sound auditory oddball task: frequent Standard tones (70%), rare Target tones (15%) and rare Novel sounds (15%). The researchers measured two components, the N100 and the P300.
Experiment 1 compared four EEG systems:
| System | Participants | Electrodes | Sampling rate | Reference (offline) |
Recording setting |
| BioSemi Active II |
10 | 64, gel, wired | 1,024 Hz | Linked mastoids |
Shielded lab |
| Smarting (mBrainTrain) |
9 | 24, semi-dry (saline sponge), wireless |
250 Hz | Linked mastoids |
Same shielded lab and audio setup |
| DSI-24 (Wearable Sensing) |
11 | 21, dry, wireless |
300 Hz | Linked ear lobes |
Non-shielded classroom |
| EPOC X (Emotiv) |
7 | 14, semi-dry (saline sponge), wireless |
128 Hz | Linked mastoids |
High-school classroom, ages 14–15 |
All data were re-referenced offline before analysis, as listed above. Recording references differed by system (see Table 1 of the paper).
Each system was tested with different participants, and the recording environments differed. The authors caution that the study is not a head-to-head hardware comparison. For Smarting, the most informative comparison is with BioSemi, because both were recorded in the same room with the same stimulus delivery.
Experiment 2 was a test–retest reliability study with 38 participants, recorded with BioSemi only.
How did Smarting mobile EEG compare with the laboratory wired system?
Individual-level detection (spatial clustering). This method tests for a response across electrodes. Smarting detected responses in 34 of 36 subject-level tests (94%). BioSemi detected them in 34 of 40 (85%).

The detection precision results for spatial clustering. Total accuracy of Semi-dry 24ch smarting mobi was 34 out of 36 ERP compnents (94%), versus 34 out of 40 components for Biosemi gel system (85%). The detection was done for 4 components and for 9 subjects for Smarting mobi (10 subjects for Biosemi system). See the table for detailed results
The study did not test these detection rates statistically between systems, and one participant changes a percentage by 10-11 points. The fair reading is that Smarting’s individual-level detection was at least comparable to the lab reference.

ERPs for all participants recorded with the in Smarting system. (A)ERPs in response to the three stimuli: Standard, target, and novel sounds, shown from electrode ‘Fz’. Shaded areas around the curves represent the standard error of the mean across trials. Time t = 0 represents our best estimation of the onset of the stimulus, as determined based on direct audio-recordings synchronized to the EEG recordings via LSL (see Methods). Horizontal black lines indicate the time points in response to each stimulus where the ERP differs significantly from zero (p< 0.05, temporal-cluster corrected). Participants (S1S9) are ordered in descending order based on their typicality score to tandard tones, which is represented by the r-value in each panel index. (B)Topographical maps showing the scalp distribution of the N100 (top) and P300 (bottom) responses, in a time window ±10 ms around the group-level peak latencies. Black dots indicate electrodes where the response differed significantly from zero (one-sided p < 0.05, spatial-cluster corrected).
Single-electrode detection (temporal clustering). This method tests one electrode across the whole epoch. Smarting detected the N100 in 78% of participants for Standard sounds, 100% for Target sounds and 89% for Novel sounds, versus 70% for each with BioSemi. P300 detection with this method was 44% for Smarting, the lowest of the four systems (see limitations below).

The detection precision results for temporal clustering. Total accuracy of Semi-dry 24ch smarting mobi was 28 out of 36 ERP compnents (78%), versus 29 out of 40 components for Biosemi gel system (73%). The detection was done for 4 components and for 9 subjects for Smarting mobi (10 subjects for Biosemi system). See the table for detailed results.
Totals in both tables are our sums of the paper’s per-measure results (Tables 5 and 6); the paper does not report totals.
Similarity to the lab reference. Smarting’s group ERPs correlated with BioSemi’s at r = 0.73 for Standard sounds and r =0.87 for Novel sounds. For Target sounds the correlation was r = 0.41. The authors found Target responses weaker across systems generally.
Waveform typicality. Typicality measures how closely each individual’s waveform resembles the group pattern. Smarting’s typicality did not differ significantly from BioSemi’s (β = 0.08, p = 0.144).
mbt comments on the paper
- The samples are small. Smarting was tested with 9 participants. The results support feasibility, not precise performance estimates.
- Different electrodes types are used. Smarting was tested only in its semi-dry configuration, compared to Biosemi’s gel electrodes.
- In semi-dry electrodes cap-fit plays a crucial role. Contrary to gel electrodes, where gel takes care of the electrode-skin contact, in semi-dry setup the cap fit is crucial, especially if the recording is rereferenced to mastoids. Because the mastoid electrode lie in the side of the head, if the cap does not fit tightly it may introduce unnecessary noise.
- Trial rejection was higher. 7.5% of Smarting trials were rejected, versus 4.7% for BioSemi and 1.8% for DSI-24. Rejection thresholds were set per system, so these rates are not directly comparable. However, Biosemi and Smarting had the same rejection threshold (25 standard deviations), despite the electrode type difference. Therefore, the rejection rate for Smarting semi-dry might be a bit too high. Other systems had softer rejection thresholds (35 and 40 standard deviations).
- The P300 temporal result: Smarting had relatively low detection rate only for temporal target P300 clustering. It was detected in 4 out of 9 participants with single-electrode temporal clustering, compared with 8 out of 9 with spatial clustering. We speculate that this might be due to above mentioned reasons: sami-dry electrodes influenced by the cap fit and hard rejection thresholds.
What should researchers take from this?
- Mobile EEG can support individual-level ERP work. In this study, a 24-channel semi-dry wireless system matched a 64-channel wired gel system on individual-level detection when both were recorded in the same room.
- Smarting reliably detects infiviual ERP responses. However, in this study, spatial clustering found individual responses more reliably than single-electrode tests.
- Smarting mobile EEG systems show very high time alignment compared to Biosemi setups.
- Novel sounds gave the most robust individual markers. Target responses were the least consistent across systems
Frequently asked questions
- Is Smarting research-grade EEG? In Korisky et al. (2026), independent researchers at Bar-Ilan University and Stanford identified Smarting as a research-grade mobile system. Tested with semi-dry saline-sponge electrodes, its individual-level ERP detection was at least comparable to a wired, gel-based 64-channel BioSemi system recorded in the same lab.
- Can semi-dry EEG match gel-based lab EEG for ERPs? In this study, semi-dry Smarting detected individual N100 and P300 responses in 34 of 36 subject-level tests, compared with 34 of 40 for a wired, gel-based BioSemi system in the same lab. So, the Smarting wireless EEG system, even with the semi-dry electrodes, showed slightly higher individual ERP accuracy than Biosemi gel laboratory EEG system
- Is this study a head-to-head comparison of EEG systems? Partly. Each system was tested with different participants, but Smarting and Biosemi systems are tested in the same environment. The authors still caution against reading the results as a hardware comparison.